Instructions
for
SCANNING TUNNELING MICROSCOPE
(STM)
By,
Eric Wittinger

Disassembling the STM

  1. Open acoustic box
  2. Disconnect the 2 main connectors (plug and coax cable) on STM (Figure 1)
  3. Figure 1. Connectors.

  4. Take STM unit our of acoustic box (Figure 2)
  5.  

    Figure 2. STM unit and Acoustic box.

  6. Loosen course positioning screws (2) until sample holder is free. (figure 3)

Figure 3. Positioning Knobs.

  1. Remove sample holder from STM Module (Figure 4)

Figure 4. Sample holder.

Cleaning Sample

1. Supplies needed Scotch Tape.

  1. Take a strip of tape and firmly press onto the sample on the sample holder
  2. Carefully remove the tape
Changing the Tip
  1. Remove the Sample plate
  2. Remove the tip from the tip holder (Figure 5)

Figure 5. Tungsten Tip.

  1. Select a new tip
  2. Using fine tipped pliers pace ¼ of the tip into the tip holder
  3. Slightly bend the tip to one side of the holder
  4. Press the another ¼ of the tip into the holder
  5. Slightly bend the tip in the opposite direction of step 5
  6. Press the tip the rest of the way into the holder.
Reassembling the STM and Adjusting the Tip
  1. Carefully place sample holder plate into STM module (Figure 6)

Figure 6. Assembled STM Unit.

  1. Using the 2 course adjusting screws roughly adjust the sample holder to obtain a flat sample plain (See Figure 3)
  1. Use a light microscope to due fine adjusting of the tip close to the sample surface
(Figure 7)

Figure 7. Light Microscope.

  1. Place the STM module back into the acoustic box on the dampening unit (Figure 8)
  2. Figure 8. Dampening Unit.

  3. Reattach the 2 main connectors (plug and coax cable)
Powering up the up STM
  1. Turn on computer
  2. Make sure the Isolation Power supply is hooked up to power the STM (Figure 9)

Figure 9. Isolation Power Supply.

  1. Turn on the CSTM 9000 Unit I, power switch located on back of unit (Figure 10)

Figure 10. CSTM 9000 Unit.

 

 

 

 

Figure 11. Vbias, Iref, Zvoltage.

  1. Turn on High Voltage Unit IV, a large red toggle switch (Figure 12)

Figure 12. High Voltage Toggle.

Setting up Computer Program

  1. Locate VSCAN1.EXE on the computer
  2. Change directories to C:\STM to Find VSCAN1
  3. Run this program by typing VSCAN1 at prompt. C:\STM\VSCAN1
  4. Press <ENTER>
  5. Main Menu will appear with options (Figure 13)
 

Fine Adjustments of the STM

  1. Use the 2 course adjustment screws to move the tip close to the sample
  1. Reseal the acoustic box
  2. Let system settle out vibrations for approximately 5 minutes.
To View Sample
  1. Select T this will start the computer scanning the sample
  2. The computer screen will automatically change and start to display the sample scans
  3. ESC to return to the main menu
Other Modes of Scanning the sample
  1. To change the scanning method flip the output switch on the CSTM-9000 unit (Figure 13)

Figure 13. Scanning Method Switch.

Powering Down the STM
  1. Quite VSCAN1 program
  2. Open up Acoustic box and back the tip away from the sample surface
  1. Turn off the High Voltage switch on the CSTM-9000
  2. Turn off unit I (Power switch is on back of unit)