Anion
Analysis 1 - Simultaneous Determination of Chloride and Bromide by a
Fluorescence Quenching Method
Introduction
Fluorescence
spectrophotometry in its pure form is inherently a very sensitive analytical method.
Single molecules have been observed in fluorescence microscopy due to the very
low signal in the absence of fluorescent analytes (often referred to as zero
background) and the possibility of producing many photons from a single
molecule using a very bright excitation source. There are two types of
fluorometric methods used in the determination of inorganic species. Direct
methods involve the formation of a fluorescent chelate and the measurement of
its fluorescence. Indirect methods are based on the quenching of a fluorescent
indicator by the species being determined. The latter technique has been most
widely used for anion analysis. Note that the indirect method (used here) is
not a zero background method as regards the analyte.
The
quenching, a decrease in fluorescence intensity, of a fluorescent indicator
(usually an aromatic organic dye) can be described by the Stern - Volmer
equation:
F0
/ F = 1 + K [Q] (1)
where F0 and F
are the fluorescence intensity in the absence and presence of a quencher (Q), K
is the quenching constant, and [Q] is the molar concentration of the quencher.
If the value of K is known or has been determined, the concentration of the
quencher [Q] can be determined by measuring F0 and F.
If
two quenchers (QA and QB) are present and operate
independently of one another, the Stern - Volmer equation takes the form:
F0
/ F = 1 + KA [QA] + KB [QB] (2)
To determine the
concentrations of two quenchers (e.g.,
chloride and bromide), it is necessary to carry out two series of fluorescence
measurements with two fluorescent indicators (1 and 2).
(F0
/ F)1 = 1 + 1KCl [Cl -] + 1KBr
[Br -] (3)
(F0
/ F)2 = 1 + 2KCl [Cl -] + 2KBr
[Br -] (4)
The
molar concentrations of chloride and bromide in an unknown mixture can be
calculated by solving the above two simultaneous linear equations, if the
quenching constants (K) are all determined in separate calibration experiments.
In this experiment, you will use the fluorescence quenching method to analyze a
solid unknown challenge sample to determine the weight percentages of NaCl and
KBr. The fluorescent indicators quinine
and acridine are used. Be sure to
provide molecular structures of quinine and acridine in the Introduction
section of your report and discuss briefly why they are good fluorescers.
Apparatus
fluorescence cuvettes (3)
25 mL volumetric flasks (12)
100 mL volumetric flasks (3)
1 mL pipet (2)
Instrumentation (See Appendix for
Operating Instructions):
PTI Spectrofluorometer with
FELIX data system
Solutions available
(1) 1.25 M H2SO4:
pour out less than 5 mL, only 1 mL is needed.
(2) 1.0 x10-4
M quinine prepared in 1.25 M H2SO4: measure out about 15
mL in a small beaker.
(3) 1.0 x10-4
M acridine prepared in 1.25 M H2SO4: measure out about 15
mL in a small beaker.
Solutions to be prepared
(1) 3.0 x10-2 M NaCl solution:
Weigh out about 0.175 g (to 0.001 g) of NaCl (FW = 58.442) and quantitatively
transfer to a 100 mL volumetric flask, dissolve with deionized water, fill to
the mark, and mix thoroughly.
(2) 3.0 x10-2 M KBr solution:
Weigh out about 0.357 g (to 0.001 g) of KBr (FW = 119.002) and quantitatively
transfer to a 100 mL volumetric flask, dissolve with deionized water, fill to
the mark, and mix thoroughly.
(3) Solution of solid sample: Weigh out
about 0.15 g (to 0.001 g) of solid NaCl and 0.35 g (to 0.001 g) KBr and
quantitatively transfer both to a 100 mL volumetric flask, dissolve with
deionized water, fill to the mark, and mix thoroughly.
Calculate
and record the analytical concentrations for all three solutions.
Procedure
Note:
In this experiment, two fluorescent indicators are used to allow the
measurement of two anionic analytes. The fluorescence excitation and emission
spectra for the two indicators are different, so you will first obtain these
spectra and select appropriate excitation and emission wavelengths for each
dye. During the quenching measurements, significant errors may be introduced if
the wavelengths are not set reproducibly. To avoid changing the wavelengths and
introducing the associated errors, you should try to complete all of the
measurements with one indicator before switching to the other one. The text
below describes the set of operations for the first dye quinine but you will
perform an identical (except for the excitation and emission wavelengths)
procedure for acridine.
(1) Preparation
of solutions:
Using
twelve 25 mL volumetric flasks, prepare the samples for fluorescence
measurement by adding the following:
#1: 1 mL of 1.25 M H2SO4 only (the
solvent blank)
#2:
1 mL quinine only (the unquenched indicator)
#3,4,5,6: 1, 2, 3, and 4 mL
of NaCl standard + 1 mL of quinine (the NaCl standards)
#7,8,9,10: 1, 2, 3, and 4 mL
of KBr standard + 1 mL of quinine (the KBr standards)
#11:
1 mL of the solution of the solid sample + 1 mL of quinine (low unknown)
#12:
2 mL of the solution of the solid sample + 1 mL of quinine (high unknown)
Fill
all of the flasks to the mark with deionized water and mix thoroughly. Label
the flasks carefully to avoid mixing them up (they all look like clear liquids).
(2) Recording the
fluorescence excitation and emission spectra
Three cuvettes should be
used for each set of measurements: one for the solvent blank, one for
the unquenched indicator, and one for the standards or samples.
Record the fluorescence excitation
spectrum for quinine as explained in the Appendix. First you will record a
blank using the sulfuric acid solution in flask #1 and the software will store
the blank value. Next you will set the emission wavelength to 450 nm and the
excitation wavelength scan to go from 300 nm to 425 nm and you will record the
spectrum of the unquenched dye in flask #2. Then with the same solution in the
cuvette, you will switch to the emission spectrum mode and set the
excitation wavelength to the maximum from the excitation spectrum and the
emission scan to run from 375 nm to 650 nm. Record the spectrum and note the
maximum value to use for the emission wavelength.
(3) Measurement of fluorescence quenching
Set the excitation and
emission wavelengths to the maxima that you determined from the two spectra
recorded above. Measure the fluorescence intensity of the solutions described
above (in flasks 2 -12) using the time-based measurement.
Hint:
The instrument used in this experiment may or may not produce stable fluorescence
readings, because changes in the light source and detector can both change the
results. If the results appear irreproducible, you could try the following
optional procedure:
1)
use the solvent blank cuvette to autozero the reading, 2) measure the unquenched
and quenched samples in the order F0 F F0 to determine
the stability of the measurements; if there is a significant difference between
the first and second unquenched readings, you may need to probe the variability
by 3) doing the measurement sequence in five steps, alternatively measuring the
fluorescence of the unquenched indicator (F0), and the fluorescence
of the standard or sample (F) in the order F0 F F0 F F0.
4) If you do it this way, you can use the standard deviation of the three F0
values for a given dye as a measure of the precision in placing the
cuvette in the instrument.
Whether you use the
procedure above or not, each set of F0 and F measurements can be
used to determine the short term precision in the measurement by calculating
the standard deviation of each 30 second data set. You should comment in your
Discussion section on which of these measures of uncertainty is the most
applicable in determining the overall uncertainty of the measurement used here.
Acridine spectra and fluorescence
quenching measurments
Repeat the above procedure
for the measurements with acridine. For the excitation spectrum, start with an
emission wavelength of 488 nm.
Report: In preparing the Partial
Report describing this experiment, you should consider/complete/discuss the
following:
(1) Fluorescence spectra:
Include two figures showing
the excitation and emission spectra (combined on one plot) for the two
fluorescent dyes. Identify the maxima that you selected for the quenching
measurements on the graphs.
(2) Quenching constants:
Tabulate all data and plot
the quenching ratio (F0 /F) vs.
concentration of chloride and bromide for quinine and acridine (four
calibration sets). You can usually put two plots on a graph, if they are
distinct (i.e., if the points and
lines dont overlap). Derive the linear equations and calculate the four
quenching constants and their uncertainties. Be sure to briefly explain how you
obtained the values and uncertainties in the Discussion section of your report
and to state the quenching constants you obtain in both the Results and
Abstract sections of your report.
(3) Chloride and bromide concentrations in
the challenge samples:
Calculate the concentrations
of chloride and bromide in the unknown challenge solutions using equations 3
and 4. Comment on whether the concentrations for the low unknown and high
unknown samples make sense, based on how they were made. The proper way to
obtain an uncertainty for the unknown concentrations combines the uncertainties
in the quenching constants with the uncertainty in the F0/F ratio as
determined from the standard deviations of the F0 measurements. Be
sure to state these values in the Results section of your report and explain
clearly but briefly how you obtained the uncertainty in your Discussion
section.
(4) Chloride and bromide in the solid
sample:
Calculate the weight
percentages of NaCl and KBr in your solid sample using the concentrations
determined above. Since the solid sample contains only NaCl and KBr, we have an
additional equation: mass of sample = mass of NaCl + mass of KBr. You can thus
comment on the accuracy of the technique in determining the total anion
concentration and the specific recoveries/accuracies for chloride and bromide
(although these are clearly inter-related). A summary statement should be
included in the Abstract.
Revised 2012-1-11 - DBA