Air Quality 2 - Determination of Cd and Pb by Anodic Stripping Voltammetry

 

Introduction:

 

In the first experiment of this set, you measured the concentration of benzene in gasoline, and it is a reasonable assumption that some fraction of it ends up in the air we breathe. Benzene is one of the air toxics that were considered in a recent assessment conducted by the Oregon Department of Environmental Quality (for further information, go to http://www.deq.state.or.us/aq/toxics/). Another air toxic (chemical compound in air that can lead to public health problems) that was identified as a possible concern is cadmium, a metal that is generally found in the ionic state (Cd2+) in the aerosol phase. Aerosols are a mixture of air and particles that can be solid or liquid, but in this case, the cadmium is attached to micrometer-sized (or smaller) solid particles that can remain suspended in the air for hours to days in the urban area. For this experiment, we will use a filter-sampling apparatus to collect aerosol samples that we will analyze by anodic stripping voltammetry. We will also determine lead in the aerosol samples, since it is another known metal air toxic (neurotoxin) that can come from industrial sources, or re-suspended road dust (left over from the days when lead was added to gasoline). Finally, we will measure the cadmium and lead in a sample of drinking water, for general interest’s sake. 

 

This electrochemical experiment uses the principle of voltammetry, where the voltage of a three electrode cell is controlled to produce an oxidation or reduction and the current transferred in that process is measured. In many voltammetric methods, the current is directly proportional to the concentration of the analyte in solution. Stripping methods attain far greater sensitivity (lower limit of detection) by pre-concentrating ions from solution first (called the deposition or plating step) and then measuring the current when they are re-oxidized from the surface of the electrode or from a thin film on the surface. In principle, all of the ions from a solution could be deposited onto the surface of the electrode, allowing each analyte molecule to contribute to the signal (but this is impractical because it would require a very long deposition time with vigorous stirring). 

 

Voltammetric methods are described by the voltage vs. time waveform (see Figure 1) used for the experiment and the anticipated current vs. time response.

http://web.pdx.edu/~atkinsdb/teach/427/Expt-ASVAerosol_files/image002.jpg

Figure 1 the voltage vs. time waveform for anodic stripping voltammetry.

 

Typically these results are combined and presented as a plot of current vs. voltage, known as a voltammogram (see Figure 2). In Figure 1, the deposition step is a constant large negative voltage (in voltammetry, the x and y axes are often backwards from usual, with increasingly negative voltages going up or to the right*) that reduces all of the metal ions in solution to metal atoms. The metal atoms either plate onto the electrode surface or are dissolved in a thin mercury film (many metals form stable amalgams with mercury) leaving them available for re-oxidation. As the voltage is rapidly scanned toward more positive values during the stripping step, individual ions strip back off the electrode at their oxidation potential resulting in current peaks for each of the metals (Figure 2). This allows multiple metals (usually up to three or four) to be determined in a single analysis, as long as their concentrations are similar (so they don’t require very different deposition times). {The voltage waveform that is actually used in this experiment is a little more complicated, but contains the essential features in Figure 1 – to document the actual waveform, double click on the CH427PbCdexpt run configuration icon at some point during the experiment and either take notes on or print out the figure in the top middle. Be careful that you don’t change anything – it is best to click “cancel” when you are finished.}

 

http://t1.gstatic.com/images?q=tbn:ANd9GcTLu9tDW1I3YmFb3Ab659lwLtWdnzP5h35ik8_4JdYGvCH9tfTFpekWNTJ0eg

Figure 2 an example of an anodic stripping voltammogram for a mixture of metals. Peak height or area can be used for quantitation of the various elements.

 

*The presentation of the voltage waveform and voltammogram here is similar to that in Skoog, et al. but the VAS software uses the other convention, where increasingly negative voltages are down and to the left, as you would normally do when making a plot.

 

Apparatus

1 - 20 μL automatic pipettor (1)

20 - 200 μL automatic pipettor (1)

1.00 mL vol. pipet (1)

10.00 mL vol. pipet (2)

10 mL graduated cylinder

ASV sample cups

Ultrasonic bath

Aerosol filter sampler

 

Instrumentation (See Appendices for Operating Instructions)

Cogent/Modern Water PDV 6000 plus and VAS software

 

Solutions to be prepared (or obtained)

Electrolyte and standard kit from Cogent/Modern Water, Inc. If necessary, prepare CLAC electrolyte by dissolving one CLAC Electrolyte Concentrate into one Electrolyte Diluent A (cap and shake until concentrate is fully dissolved).

 

Procedure

(1) Collection and Extraction of Aerosol Filter Sample:

Do this part of the procedure well before you do the laboratory experiment.

·         Place a filter into the plastic Buchner funnel and support the funnel so that it is up off the ground and away from walls and other surfaces.

·         Turn on the pump and adjust the flow to 20 lpm.

·         Record the start and stop times for the sample and any other key variables (proximity to traffic, meteorology, etc.)

·         Remove the filter and wrap it in aluminum foil.

 

If you are going to analyze a cigarette sample (tobacco or ash) or a smoke filter sample, you will follow the extraction procedure below to prepare the sample for analysis.

 

Do this part of the procedure in the fume hood near the ASV at the beginning of the lab period when you will be analyzing the samples using ASV.

·         Prepare a 10 mL graduated cylinder with ~ 1.5 M nitric acid: fill to 9 mL with DI water and carefully add 1 mL of concentrated HNO3.

·         Place the filter in the bottom of a 100 mL beaker and add the nitric acid solution.

·         Carefully place the beaker in the ultrasonic bath with just enough water to avoid floating the beaker and agitate for 15 minutes.

·         Remove the filter from the beaker using a stir rod and neutralize it in a small beaker half-filled with sodium bicarbonate before placing it in the solid waste container.

·         Put the beaker near the center (inward) of the hood and carefully add 3 solid sodium hydroxide pellets. Swirl until the pellets dissolve.

·         Cover the beaker with parafilm until you are ready to do the analysis.

 

(2) ASV Measurement of cadmium and lead in water sample using Standard Comparison (SC): Important! Do not turn on the PDV6000plus without placing the electrodes in a sample cup containing a suitable electrolyte. The resulting voltages can damage the Working Electrode. See the Appendix for further information. 

The World Health Organization has recommended water quality standards of no more than 3 ppbv cadmium and 10 ppbv lead. We do not expect your drinking water to contain anything close to those levels, so we will conduct a screening (rather than a proper determination) on this sample. Because your water sample has very little electrolyte (and ASV requires a supporting electrolyte) we will compromise between providing an electrolyte and diluting the sample with it – i.e., we will add of 10.00 mL of the water samples to 10.00 mL of the CLAC electrolyte in a pre-labeled ASV measurement cup. The Standard Comparison standard will be 20.00 mL of CLAC electrolyte with 40 µL of a 10 ppm lead/cadmium standard (to produce a solution with 20 ppbv Cd and 20 ppbv Pb). Note that your values for the water samples may fall outside of the SC confidence range and that will be indicated by a “<” in front of the number or just a zero reported concentration.

 

(3) Prepare a set of three challenge “unknown” samples: 1) a nominal 10 ppbv Cd “unknown”, 2) a nominal 10 ppbv Pb “unknown”, and 3) a solution nominally containing both 10 ppbv Cd and 10 ppbv Pb. These solutions are to be produced by appropriate dilutions of the 20 ppm Cd and Pb standards in the CLAC electrolyte to produce 20 mL of solution in pre-labeled ASV measurement cups.

 

Remember to create a new sample file between each measurement (SC vs. SA and challenge vs. real unknowns)

 

(4) ASV Measurement of cadmium and lead in challenge and aerosol samples using Standard Addition (SA):

The Appendix assumes that you are analyzing the aerosol sample first, but you may find it easier to analyze the challenge “unknowns” first, since you know their concentrations and thus the logical size of the spikes. (And it is encouraging to start with solutions that actually give results.) If so, remember to rinse the cell at least three times between each sample and the next blank (there isn’t a standard run in SA, so you don’t have that extra rinse step). For the single metal challenge unknown samples (Cd or Pb only) it obviously makes more sense to use the single metal standards (which are 20 ppm) to do the spikes.

 

The concentrations of lead and cadmium in atmospheric aerosol are highly variable and sometimes very low. So the first thing we will do to improve our chances of seeing the metals is increase the deposition time to 5 minutes. Double click on the CH427PbCdexpt run configuration icon and change the deposition time from 180 seconds to 300 seconds and set the configuration as active. (We don’t do this for all the samples because it takes too long. Make sure you change back if you run the challenge samples after the aerosol sample.) The beauty of the standard addition procedure is that you run the sample first (after a blank) and if there is no measureable concentration of one or both metals, it will be readily apparent. If this is the case, you should still do two 10 ppbv spikes to establish the limit of detection for the method. You will then be able to report the metal(s) as a non-detect and provide an upper limit for the concentration. Optimally, spikes should be added that are individually smaller than the concentration (~ 50% or more) but enough spikes should be used to more than double the original concentration, clearly establishing the analytical sensitivity (the slope of the - hopefully linear - SA curve). If the lead and cadmium concentrations are vastly different, it may be easier to do the SA procedure for each separately, using another 1 mL of the extractate to prepare another sample. Be sure that you take careful notes on what you actually do, and document your procedure in the Experimental section of your report.

 

Report: In preparing your Final Report, you should consider/complete/discuss the following (not necessarily in this order):

 

(1) Comment on the voltammograms and whether you see any shifts in the peak positions for either metal. These shifts are sometimes seen with concentration (for some metals) or more commonly with “real” samples, where the metals may be partially complexed.

 

(2) Present and comment on the appearance of the standard addition curves. If they are not linear, estimate a corrected value(s) for the unknown(s) based either on a smaller number of spikes or a non-linear fit to the data (note the peak heights and areas are provided in the Results item). 

 

(3) Comment on the accuracy of the SA method, based on the challenge unknown results. What is the relative accuracy of this version of ASV? Is this acceptable for environmental applications (in your opinion)?

 

(4) If any of your samples (water or aerosol) contains measureable levels of lead or cadmium, calculate the relevant concentration. (For any values that are below the measurement threshold, state the result as a non-detect with an estimated LoD.) For water samples, simply apply any relevant dilution factors. For the aerosol samples, you can report micrograms per cubic meter of air, based on the total metal mass in the sample and the volumetric air flow through the filter (20 L/min) multiplied by the total sampling time. Comment on the values, but don’t be too worried for your health, there is ample opportunity for contamination in this procedure.

 

Revised 2014-1-21 DBA